  • Digits of Resolution - 6.5
  • Max. Reading Speed - 30.000 readings/s

The DMM1 module is designed to operate as part of information measuring systems based on the VXI bus as a digital multimeter. The module is a device on the basis of registers and has the dimensions C-1.

The module performs the measurement of direct and alternating voltage, direct and alternating current, resistance, frequency.

When measuring DC voltage in the ranges of 100 mV, 1 V and 10 V; input impedance can be selected 10 MOhm or 10 GOhm. In the ranges of 100 V and 400 V, the input resistance is fixed and equal to 10 MOhm ±1%.

Active resistance measurement is performed with the possibility of choosing one of the measurement schemes: 2-wire or 4-wire. The maximum polling current does not exceed 1 mA. The output of the ohmmeter reference current source is protected from applying an external voltage with an amplitude of ±400 V.

The current is measured using the built-in shunt whose resistance is 0.1 Ohm for the measuring ranges 1 A and 3 A and 5.1 Ohms for the ranges 10 mA and 100 mA. Protection of the input of the ammeter against excessive current is carried out using a fuse, which is accessed from the front panel of the module.

All measurement inputs are galvanically isolated from the housing. Voltage of galvanic isolation is not less than 500 V.

The module contains an internal memory that allows storing up to 4K measurements once. The organization of memory allows reading from it without interrupting the measurement process, which allows an endless measurement mode.

The module has the ability to synchronize hardware with external devices via the TTLTRG line or the “Start” and “End of measurement” connectors located on the front panel.

The module has an automatic zero correction mode.

The module has an automatic range detection mode.

The capacity of the module is 6.5 characters (with an aperture of 1, 10 and 100 periods of the supply network) Maximum measurement speed - 30,000 measurements / sec (with a digit capacity of 4.5 characters)

Input impedance

  • 10 MOhm or 10 GOhm (in the range of 100 mV, 1 V, 10 V),
  • 10 MOhm ± 1% (in the range of 100 V, 400 V)
2-wire or 4-wire resistance measurement circuit
Maximum polling current, no more than 1 mA

Built-in Shunt Resistance

  • 0.1 Ohm (when measured in the range of 1 A, 3 A),
  • 5.1 Ohm (with a change in the range of 10 mA, 100 mA)
Voltage of galvanic isolation of measuring inputs from the housing, not less than 500 V

Ranges of measurement of direct voltage, direct current and resistance

Measured function Range Error ±(A+B)%
A B
Constant pressure ( U ) 100.0000 mV 0.0040 0.0035
1.000000 V 0.0030 0.0007
10.00000 V 0.0020 0.0005
100.0000 V 0.0035 0.0006
400.0000 V 0.0035 0.0030
 DC ( I )
10.00000 mA 0.05 0.01
100.0000 mA 0.040 0.005
1.000000 A 0.130 0.010
3.000000 A 0.720 0.020
Resistance ( R ) 100.0000 Ohm 0.008 0.004
1.000000 kOhm 0.008 0.001
10.00000 kOhm 0.008 0.001
100.0000 kOhm 0.008 0.001
1.000000 MOhm 0.008 0.001
10.00000 MOhm 0.035 0.001
100.0000 MOhm 0.800 0.010
     Notes
  1. A - percent of the measured value, B - percent of the range
  2. The upper limit of the range is: 100% of the nominal value of the senior range and 120% of the nominal value for the remaining ranges.
  3. The error is given for an aperture of 10 and 100 periods of the supply network.

Ranges of measurement of alternating voltage and alternating current

Range
Frequency Error ±(A + B)%
A B
AC voltage
100mV-300V 3Hz-5Hz 1.00 0.03
100mV-300V 5Hz-10Hz 0.35 0.03
100mV-300V 10Hz-20kHz 0.05 0.03
100mV-300V 20kHz-50kHz 0.11 0.05
100mV-300V 50kHz-100kHz 0.60 0.08
100mV-300V 100kHz-300kHz 5.00 0.50
AC
1A, 3A 3Hz-5Hz 1.5 0.06
5Hz-10Hz 0.95 0.06
10Hz-1kHz 0.75 0.06
1kHz-50kHz 1.00 0.06