MOSC5

MOSC5
Order
Item article:
Order
  • 2 channels
  • 14 bit ADC
  • 128 mtsch / channel memory
  • Bandwidth 200 MHz
  • Signal measurement from -150 to 150 V

MOSC5 is designed to convert instantaneous voltage values of an electric signal into a digital code, measure instantaneous voltage values, measure time intervals between two instantaneous voltage values through two independent channels with galvanic isolation.

MOSC5 provides:

  • digital-to-analog conversion of the instantaneous voltage values ​​of the input signals through two channels with a given sampling rate and recording the received measurement information in the memory installed on the mezzanine.
  • the ability to programmatically connect to the object of control and disconnect from the control object of the input signal lines of the measuring channels
  • the ability to set launch parameters: fountain of the trigger signal for the "internal" trigger mode; a trigger signal slice for the “internal” trigger mode; measurement time preceding the appearance of a sync event;
  • the ability to select the following triggering events: at the command of the program (“program” launch mode); achievement by the input signal of the set trigger threshold on one (selected) of the two channels (“internal” trigger mode); the input signal reaches the set trigger threshold on the analog sync input (connector "CB" on the front panel of the mezzanine);
  • software installation of the following modes of DC channel operation: “open input” (absence of dividing capacity at the input); "Closed entrance" (presence at the input of the separation tank)

MOSC5 channels are galvanically isolated from the mezzanine case and control signals.

Measurement of instantaneous voltage values at the inputs of each channel
Range designation
«± 0,05 V» «± 0,1 V» «± 0,25 V» «± 0,5 V» «± 1 V» «± 2,5 V»
Input voltage swing, V
-0,05 to  +0,05     -0,10 to +0,10     -0,25 to +0,25     -0,50 to +0,50
-1,00 to +1,00     -2,50 to +2,50    

Range designation
«± 5 V» «± 10 V» «± 25 V» «± 50 V» «± 100 V» «± 150 V»
Input voltage swing, V
-5,00 to +5,00 -10,00 to +10,00 -20,00 to +20,00 -50,00 to +50,00 -100,00 to +100,00 -150,00 to +150,00
Minimum 3 dB bandwidth of each channel: 50 MHz The number of bits of the ADC in each channel - 14
Software setting of input resistance values for each channel:
50 Ohm ±2% for range 1 MOhm ±1% for range    
«± 0,05 V» «± 0,05 V»
«± 0,1 V» «± 0,1 V»
«± 0,25 V» «± 0,25 V»
«± 0,5 V» «± 0,5 V»
«± 1 V» «± 1 V»
«± 2,5 V» «± 2,5 V»
«± 5 V» «± 5 V»
  «± 10 V»
  «± 25 V»
  «± 50 V»
  «± 100 V»
  «± 150 V»
The signal bias voltage for each channel is set within the selected measurement range. The margin of error of the bias voltage setting, reduced to the value of the upper limit of the measurement range, is ± 0.6%. Resolution resolution of bias voltage:
  • 100 μV in the range «± 0,05 V»;
  • 200 μV in the range «± 0,1 V»;
  • 500 μV in the range «± 0,25 V»;
  • 1 mV in the range «± 0,5 V»;
  • 2 mV in the range «± 1 V»;
  • 5 mV in the range «± 2,5 V»;
  • 10 mV in the range «± 5 V»;
  • 20 mV in the range «± 10 V»;
  • 50 mV in the range «± 25 V»;
  • 100 mV in the range «± 50 V»;
  • 200 mV in the range «± 100 V»;
  • 300 mV in the range «± 150 V»
The limits of the upper limit of the range of permissible error in measuring instantaneous voltage values for each channel are ± 0.6%
The maximum non-destructive voltage supplied to the inputs of the mezzanine channels at a time of continuous exposure not exceeding 1 min, no more than:
  • ± 7.5 V at an input impedance of 50 Ohm;
  • ± 200 V at an input impedance of 1 MOhm
Limits of permissible absolute error of measurement of time intervals in the measurement range from 5.12 μs to 13743.686 s

± [0,0000025Tx + Td],

where Tx – measured time interval; Td - sampling period
Signal measurement with a maximum sampling rate of 200 MHz. Setting the sampling period (TD): 5 ns and from 10 to 655350 ns in 10 ns increments Software setting of the same measurement time for both channels in the range from 5.12 μs to 13743.686 seconds in increments (Td × 1024), where Td is the sampling period. Within the range, the measurement time is determined by the formula
N × Td,
where N is the number of samples written to RAM during one measurement cycle; Td - sampling period
RAM to store conversion results for each channel
The trigger threshold setting range in the “internal” trigger mode corresponds to the selected measurement range. The resolution increment of the trigger threshold is equal to the value <value of the upper limit of the measurement range, V> / 128 The measurement time preceding the appearance of the sync event is set in the range from 16Td to (T - 16Td) seconds in increments of 16Td seconds, where Td is the sampling period, T is the measurement time
The signal generated from the internal trigger source has the shape of a rectangular pulse, the parameters of which correspond to the logic levels of the TTL type IC:
  • logical zero level - from 0 to 0.8 V;
  • logical unit level - from 2.4 to 5 V
Software setting of parameters at start-up from a signal supplied to the CB connector:
  • trigger level - from minus 5 to plus 5 V in 0.1 V increments;
  • trigger event - slice / edge of the signal.
Software selection of reference frequency sources:
  • internal generator;
  • an external generator, from which a signal with a frequency of 10 MHz should be supplied to the connector “VCh VX” mezzanine. The input signal level must correspond to the logic levels of the TTL type IC (log. 0 - from 0 to 0.8 V; log. 1 - from 2.4 to 5.5 V)
At the command of the program, a signal is generated from the internal reference frequency generator. Signal parameters of the internal reference frequency generator:
  • frequency - 10 MHz;
  • the levels of the generated signal correspond to the logical levels of the TTL type IC (log. 0 - from 0 to 0.8 V; log. 1 - from 2.4 to 5.5 V)
Rise time in each channel no more than 7 ns The electrical strength of the galvanic isolation is not less than 200 V. The resistance of the galvanic isolation is not less than 20 MOhm
MOSC5