The IS4 precision digital resistance meter is designed for use in the upgraded version of the quality control system for mounting and testing  cable networks TEST-9110. The IS4 module can also be successfully used in other instrumentation systems built on the basis of the open international standard VXI.

Key Features:

  • A new mode “Checking the quality of installation” has been introduced.
  • In all modes, the “Autorange” mode has been introduced, due to which the data is supplemented by higher digits, which will contain information about the range to which this data belongs.
  • The amount of RAM is reduced to 1K measurements.
  • 3 additions were made to the operating panel of the IS4 module: the formation of direct current strength, the measurement of differential DC voltage and the measurement of circuit resistance.

The module provides:

  • reproduction of a direct current voltage;
  • DC voltage measurement;
  • DC current measurement;
  • reproduction of voltage of alternating current;
  • measuring the rms value of the AC voltage;
  • measuring the rms value of the AC current;
  • DC resistance measurement;
  • measurement of circuit resistance at a given value of the measurement current;
  • measurement of insulation resistance at a given value of the test DC voltage;
  • checking the isolation at a given value of the test DC voltage;
  • checking the isolation at a given rms value of the test AC voltage;
  • measurement of electrical capacitance;
  • Integrity check p-n junction.

Reproduction of a direct current voltage of positive or negative polarity:
  • from 4 to 100 V - range "100 V";
  • from 100 to 1050 V - range "1000 V";
  • from -4 to -100 V - range "-100 V";
  • from -100 to -1050 V - range "-1000 V"
Resolution resolution of DC voltage in the ranges:
  • 0.5 V in the ranges of "100 V" and "-100 V";
  • 1,0 V in the range of "1000 V" and "-1000 V"
Limits of permissible relative error in setting the DC voltage:
  • ± 2.5% - for values from 4 to 9.5 V and from -4 to -9.5 V;
  • ± 1% - for values from 10 to 1050 V and from -10 to -1050 V
The range of ripples of the output DC voltage is not more than (0.02 Uout + 0.1) V, where Uout is the output voltage
The DC voltage source contains an overcurrent protection circuit The threshold values of the trip current of the overload protection circuit for direct current of positive and negative polarity are set programmatically:
  • from 0.1 to 1 mA - the range of "1 mA";
  • from 1 to 10 mA - range "10 mA"
Discretization of setting the threshold value of the trip current of the DC overload protection circuit 0.1 mA Upon reaching a predetermined threshold value of the trip current of the DC overload protection circuit, the DC voltage source is turned off
The limits of the tolerance, reduced to the upper limit of the range, of the error in setting the threshold value of the trip current of the overload protection circuit in each range of ± 5% Software setting of the rise and fall time of the test voltage in the range from 0.01 to 1.2 s at the rise and fall rate of not more than 10000 V / s. Step of setting the rise time and removal of the test voltage of 0.01 s
Software setting of the test voltage holding time in the range from 0 to 655 s. 0.01 s test voltage retention time step Installation of a test voltage program command with a specified rise time and a specified voltage value, as well as removal of a set test voltage by a program command
Measurement of DC voltage of positive and negative polarity in the following ranges:
  • from -10 to 10 V - range "10 V";
  • from -100 to 100 V - range "100 V";
  • from -700 to 700 V - “700 V” range
The DC voltage is measured with the following integration times:
  • 2 ms;
  • 20 ms (one period of the supply network 50 Hz);
  • 200 ms (ten periods of power supply 50 Hz)
The permissible limits, reduced to the upper limit of the range, of the error in the measurement of DC voltage in each range:
  • ± 0.5% at integration time equal to one or ten periods of the supply network;
  • ± 2.0% with an integration time of 2 ms
Measurement of direct current of positive and negative polarity:
  • from -100 to 100 μA - the range of "0.1 mA";
  • from -1 to 1 mA - the range of "1 mA";
  • from -10 to 10 mA - range "10 mA";
  • from -100 to 100 mA - “100 mA” range
The input resistance of the DC voltage meter is not less than 9.8 MOhm The DC current is measured with the following integration times:
  • 2 ms;
  • 20 ms (one period of the supply network 50 Hz);
  • 200 ms (ten periods of power supply 50 Hz)
The permissible limits, reduced to the upper limit of the range, of the error in the measurement of direct current in each range:
  • ± 0.5% at integration time equal to one or ten periods of the supply network;
  • ± 2.0% with an integration time of 2 ms
The maximum voltage drop across the measuring circuits of the DC current meter is not more than 0.7 V
Reproduction of a sinusoidal voltage of alternating current in the range from 100 to 750 V of mean square value
Frequency of the reproduced sinusoidal voltage of alternating current 50 Hz The limits of the permissible relative error of the frequency of the reproduced sinusoidal voltage of the alternating current ± 0.5%
The harmonic distortion coefficient of the sinusoidal voltage of the alternating current is not more than 10% Resolution of setting the rms value of AC voltage 1 V
Limits of permissible relative error in setting the rms value of the AC voltage ± 5% Software setting of rise, hold and removal times of test AC voltage:
  • the rise and fall times are set equal from 0 to 1.2 s. Installation step 0.1 s;
  • retention time - from 0.02 to 655 s. Installation step 0.02 s
The AC voltage source contains an overcurrent protection circuit The threshold rms currents of the AC overload protection circuit are set programmatically in the range from 0.1 to 7 mA
Discreteness of setting the threshold rms value of the trip current of the overload protection circuit for alternating current 0.1 mA When the set threshold value of the tripping current of the AC overload protection circuit is reached, the AC voltage source is turned off
Limits of tolerance, reduced to the upper limit of the range, of the error in setting the rms threshold current ± 10% Measuring the rms value of the AC voltage in the ranges:
  • from 0 to 10 V - range "10 V";
  • from 0 to 100 V - range "100 V";
  • from 0 to 700 V - range "700 V"
The limits of permissible, reduced to the upper limit of the range, measurement error of the rms value of the alternating current voltage in the frequency range from 50 Hz to 1 kHz with an integration time of at least one period of the supply network ± 1% The input resistance of the AC voltage meter is not less than 0.98 MOhm
The time of one measurement of the rms value of the AC voltage is not more than 100 ms
Measuring the rms value of the AC current in the range from 0 to 100 mA The limits of the permissible, reduced to the upper limit of the range, measurement error of the rms value of the AC current in the frequency range from 50 Hz to 1 kHz with an integration time of at least one period of the mains ± 5%
The maximum voltage drop across the measuring circuits of the AC meter is not more than 0.7 V rms
The time of one measurement of the rms value of the AC current is not more than 100 ms DC resistance measurement using a 4-wire measurement circuit in the ranges:
  • from 0 to 10 Ohms - the range of "10 Ohm";
  • from 0 to 100 Ohms - the range is “100 Ohm”;
  • from 0 to 1 kOhm - the range is “1 kOhm”;
  • from 0 to 10 kOhm - the range is “10 kOhm”;
  • from 0 to 100 kOhm - the range is “100 kOhm”;
  • from 0 to 1 MOhm - the range is “1 MOhm”;
  • from 0 to 10 MOhm - the range of 10 MOhm;
  • from 0 to 100 MOhm - range "100 MOhm"
DC resistance is measured at the following programmable integration time values:
  • 2 ms;
  • 20 ms (one period of the supply network 50 Hz);
  • 200 ms (ten periods of power supply 50 Hz)
The permissible limits, reduced to the upper limit of the range, of the error in measuring DC resistance:
  • with an integration time of 2 ms:
    • ± 0.5% in the ranges of “10 Ohm”, “100 Ohm”, “1 kOhm”, “10 kOhm”, “100 kOhm” and “1 MOhm”;
    • ± 1% in the range of "10 MOhm";
    • ± 5% in the range of "100 MOhm";
  • when the integration time is equal to one or ten periods of the supply network:
    • ± 0.1% in the range of "10 Ohm";
    • ± 0.03% in the range of "100 Ohm";
    • ± 0.02% in the ranges of “1 kOhm”, “10 kOhm”, “100 kOhm” and “1 MOhm”;
    • ± 0.1% in the range of "10 MOhm";
    • ± 1.5% in the range of "100 MOhm"
When measuring direct current resistance, the value of direct current supplied to the measured resistance does not exceed 5.5 mA
The maximum value of the voltage drop across the measured resistance when the measuring current flows through it is not more than:
  • 2 V in the ranges of “10 Ohm”, “100 Ohm”, “1 kOhm”, “10 kOhm” and “100 kOhm”;
  • 8 V in the ranges of “1 MOhm”, “10 MOhm” and “100 MOhm”
Measurement of circuit resistance at a given value of the measurement current according to a 4-wire circuit
In the circuit resistance measurement mode, the module provides reproduction of direct current measurement in the ranges:
  • from 5 to 100 mA - the range of "100 mA";
  • from 0.1 mA to 2 A - range "2 A".
Discretization of installation of measurement current in each range of 1 mA. Limits of permissible relative error in setting the measurement current in each range ± 0.5%
In the circuit resistance measurement mode, the module provides voltage limitation on the measured circuit in the ranges:
  • from 0.5 to 30 V at a current of up to 1 A;
  • from 0.5 to 20 V at a current of 1 to 2 A.
The resolution of setting the voltage limit level on the measured circuit is 0.1 V. The permissible relative error of setting the voltage limit level on the measured circuit is ± 0.5%
  • With a measurement current of 1 A and a voltage limitation level of 30 V, the module provides a measurement of circuit resistance in the range from 0.002 to 30 Ohm.
  • With a measurement current of 2 A and a voltage limit level of 20 V, the module provides a measurement of circuit resistance in the range from 0.002 to 10 Ohm.
  • At other values ​​of the measurement current, the module provides a measurement of the circuit resistance in the range from 0.002 Ohm to Rв, where: Rв - the upper limit of the range in ohms. Rb = Uo / Ii; Uо - the value of the set level of voltage limitation on the measured circuit in volts; Ii - value of test current in A
The circuit resistance is measured with the following programmable integration time values:
  • 2 ms;
  • 20 ms (one period of the supply network 50 Hz);
  • 200 ms (ten periods of power supply 50 Hz)
The limits of permissible relative error in measuring the resistance of the circuit with an integration time of at least one period of the supply network correspond to the values

± (0,5 + 0,01 / (Iи × Rx) %, where Iи is the set value of the measurement current in A; Rx is the value of the measured resistance in Ohm

Measurement of insulation resistance in the range from 0.1 to 1000 MOhm

Limits of relative error in measuring the isolation resistance

Measured Leakage Current Range (Iу)

Intervals for leakage currents

Iун < Iу ≤ Iув

Permissible relative error, %

Iун, mA Iув, mA

«0,1 mA»

(Iу ≤ 0,1 mA)

0,01

0,1

± 1

0,001

0,01

± 2

0,0001

0,001

± 5

0

0,0001

± 10

«1 mA»

(Iу ≤ 1 mA)

0,1

1

± 1

0,01

0,1

± 2

0,001

0,01

± 5

0

0,001

± 10

«10 mA»

(Iу ≤ 10 mA)

1,0

10

± 1

0,1

1,0

± 2

0,01

0,1

± 5

0

0,01

± 10

Note - The leakage current is determined by the formula Iу = Uisp / Rism, where Uisp - the value of the test voltage from 4 to 1000 V; Rism - measured resistance value from 10 E + 3 to 1 E + 9 Ohm   
Checking the insulation strength with a DC voltage set programmatically in the range from 4 to 1050 V, or an AC voltage set in the range from 100 to 750 V rms Checking the insulation strength with DC voltage with a programmable breakdown current limit in the range from 1 to 10 mA in 0.5 mA steps
Checking the insulation strength with AC voltage with a software-defined breakdown current limit in the range from 1 to 10 mA rms in increments of 0.5 mA Measurement of electric capacitance in the ranges:
  • from 0.1 to 1 nF - the range of "1 nF";
  • from 1 to 10 nF - the range of "10 nF";
  • from 10 to 100 nF - the range of "100 nF";
  • from 100 to 1000 nF - the range of "1000 nF"
Limits of permissible relative measurement error of electric capacity ± 10% The time of one measurement of electric capacity is not more than 0.02 s
Check integrity pn transition. The test current is set programmatically in the range from 1 to 10 mA in increments of 1 mA The time of one integrity check of the p-n junction is not more than 0.1 s
In the modes of measuring the insulation resistance and checking the insulation strength, the module provides the choice of one of the connectors, through which the measured current "Ix" (circuit "Ix / -Ux") is supplied: the connector "X1" or the connector "Ix" Detachment of all input and output lines of the module from the contacts of the output connector by the program command
Internal connection at the command of the program of current and measuring lines to ensure the measurement of DC resistance according to a 2-wire circuit The outputs of voltage and current sources, as well as the measurement inputs of the module are galvanically isolated from the housing. The electrical strength of the galvanic isolation is not less than 1000 V. The resistance of the galvanic isolation is not less than 1 GOhm
Indication of the presence of test voltage at the output contacts of the connector. The indicator is located on the front panel of the module. Auto Range Measurement Mode
Checking the main technical characteristics in the "Self-control" mode The possibility of a functional self-test when using an external cable in the “OK disconnected” and “OK disconnected 2” modes